Code on display ERR06 Gram Precision TGI / TG

Instability in density mode

In density measurement, the weighings in air and in liquid must be stable. Fluctuation invalidates the result.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Close the draft shield (glass doors) and eliminate air currents.
  2. Move the balance to a sturdy table free of vibration.
  3. Check that the sample hangs freely in the liquid and does not touch the container.

Possible causes

  • Air currents (air conditioner, door, fan)
  • Table vibration
  • Temperature change or warm-up time not observed

Do you need service?

It resolves when the environmental conditions are corrected. If it recurs continuously in a proper environment, a service check is needed.

Models showing this code: TGI Precision, TGI Analytical, TG

Source: Gram Precision official user manual