Calibration did not start due to instability
Calibration can only be started on an empty, stable weighing pan. The balance rejects the operation because a fluctuating reading would compromise calibration accuracy.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Close the draft shield (glass doors) and eliminate air currents.
- Move the balance to a sturdy table free of vibration.
- Empty the weighing pan and press the CAL key again after the stability icon lights up.
Possible causes
- Air currents (air conditioner, door, fan)
- Table vibration
- Temperature change or warm-up time not observed
Do you need service?
It resolves when the environmental conditions are corrected. If it recurs continuously in a proper environment, a service check is needed.
Models showing this code: TGI Precision, TGI Analytical, TG
Source: Gram Precision official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.