Code on display ERR02 Gram Precision TGI / TG

Calibration did not start due to instability

Calibration can only be started on an empty, stable weighing pan. The balance rejects the operation because a fluctuating reading would compromise calibration accuracy.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Close the draft shield (glass doors) and eliminate air currents.
  2. Move the balance to a sturdy table free of vibration.
  3. Empty the weighing pan and press the CAL key again after the stability icon lights up.

Possible causes

  • Air currents (air conditioner, door, fan)
  • Table vibration
  • Temperature change or warm-up time not observed

Do you need service?

It resolves when the environmental conditions are corrected. If it recurs continuously in a proper environment, a service check is needed.

Models showing this code: TGI Precision, TGI Analytical, TG

Source: Gram Precision official user manual