Incorrect calibration weight
The weighing module did not accept the placed weight as the calibration value. Calibration is not completed if the weight is wrong or the environment contains disturbances.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Use a calibration weight of the correct value.
- Eliminate air currents and vibration in the environment.
- Repeat the calibration.
Possible causes
- Calibration weight of the wrong value
- Air drafts or vibration
- Sample holder fitted incorrectly
Do you need service?
If it persists when retried with the correct weight, a service check is needed.
Models showing this code: TH
Source: Gram Precision official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.