Code on display ERR09 Gram Precision TGI / TG

Instability in formulation mode

In formulation, each component is recorded after a stable reading. Fluctuation prevents the component from being recorded.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Close the draft shield (glass doors) and eliminate air currents.
  2. Move the balance to a sturdy table free of vibration.
  3. After adding the component, wait for the stability icon and confirm.

Possible causes

  • Air currents (air conditioner, door, fan)
  • Table vibration
  • Temperature change or warm-up time not observed

Do you need service?

It resolves when the environmental conditions are corrected. If it recurs continuously in a proper environment, a service check is needed.

Models showing this code: TGI Precision, TGI Analytical, TG

Source: Gram Precision official user manual