Calibration could not be started
The weighing module of the moisture analyzer expects a stable zero for calibration. The operation does not start if fluctuation continues.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Close the heating cover and wait for the instrument to cool down.
- Eliminate sources of air currents and vibration.
- Restart the calibration.
Possible causes
- Air drafts
- Vibration
- Instrument warm-up not completed
Do you need service?
Calibration can be started in a stable environment.
Models showing this code: TH
Source: Gram Precision official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.