Code on display ERR01 Gram Precision TGI / TG

Unstable weight after tare

The scale waits for a stable reading to complete the tare operation. If the value does not settle because of air currents or vibration in the environment, the operation is not performed.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Close the draft shield (glass doors) and eliminate air currents.
  2. Move the balance to a sturdy table free of vibration.
  3. Press the TARE key again after the container and sample have settled on the weighing pan.

Possible causes

  • Air currents (air conditioner, door, fan)
  • Table vibration
  • Temperature change or warm-up time not observed

Do you need service?

It resolves when the environmental conditions are corrected. If it recurs continuously in a proper environment, a service check is needed.

Models showing this code: TGI Precision, TGI Analytical, TG

Source: Gram Precision official user manual