Code on display ERR2 Gram Precision RK

Initial zero point exceeded

At startup the value read from the load cell is too far from the zero point set at calibration. The scale cannot obtain a reliable zero in this state, so it does not start weighing. This is usually caused by an object left on or under the weighing pan and rarely by a load cell fault.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn off the scale and remove everything from the weighing pan.
  2. Check under and around the weighing pan and clear any trapped objects.
  3. Make sure the weighing pan is seated correctly, then turn the scale on again with an empty pan.
  • Do not strike the load cell or force the weighing pan.

Possible causes

  • A load was on the weighing pan at switch-on
  • Foreign object trapped under the weighing pan
  • Load cell fault or damage

Do you need service?

If the error persists with an empty pan, the manual recommends service to check/replace the load cell.

Models showing this code: RK-3, RK-6, RK-15, RK-30

Source: Gram Precision official user manual