Code on display ERR9 Gram Precision RK

Tare range exceeded

According to the manual, the tare value must be greater than zero and less than or equal to full capacity. Out-of-range tare is not accepted.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check the entered preset tare value and enter a value smaller than the capacity.
  2. Use a lighter container.
  3. Empty the weighing pan, clear the tare and repeat the operation.

Possible causes

  • Preset tare larger than the capacity entered
  • Attempt to tare at zero or a negative value
  • Container on the weighing pan exceeds the capacity

Do you need service?

This is a setting/usage error. It is fixed with the correct tare value.

Models showing this code: RK-3, RK-6, RK-15, RK-30

Source: Gram Precision official user manual