Instability in percent mode
In percent weighing, the 100% value cannot be assigned unless the reference weight is read stably. According to the manual, ERROR 07 also appears when the reference weight is entered as less than 10 divisions.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Close the draft shield (glass doors) and eliminate air currents.
- Move the balance to a sturdy table free of vibration.
- Make sure the reference weight is greater than 10 divisions and repeat the operation.
Possible causes
- Air currents (air conditioner, door, fan)
- Table vibration
- Temperature change or warm-up time not observed
Do you need service?
It resolves when the environmental conditions are corrected. If it recurs continuously in a proper environment, a service check is needed.
Models showing this code: TGI Precision, TGI Analytical, TG
Source: Gram Precision official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.