Code on display ERR07 Gram Precision TGI / TG

Instability in percent mode

In percent weighing, the 100% value cannot be assigned unless the reference weight is read stably. According to the manual, ERROR 07 also appears when the reference weight is entered as less than 10 divisions.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Close the draft shield (glass doors) and eliminate air currents.
  2. Move the balance to a sturdy table free of vibration.
  3. Make sure the reference weight is greater than 10 divisions and repeat the operation.

Possible causes

  • Air currents (air conditioner, door, fan)
  • Table vibration
  • Temperature change or warm-up time not observed

Do you need service?

It resolves when the environmental conditions are corrected. If it recurs continuously in a proper environment, a service check is needed.

Models showing this code: TGI Precision, TGI Analytical, TG

Source: Gram Precision official user manual