Code on display -FULL- RADWAG C315 scales

Measuring range exceeded

The load on the platform is above the scale's weighing range, so no value can be displayed. Continued overload can damage the load cell.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the excess load from the platform.
  2. Weigh the load in portions that fit within the capacity.
  • Do not place a load above capacity or an impact load on the platform.

Possible causes

  • Sample heavier than capacity
  • Pressure or impact on the weighing pan

Do you need service?

If FULL is also shown with an empty pan, the measuring system must be checked by technical service.

Source: RADWAG official user manual · Open manufacturer document