Code on display -null- RADWAG C315 scales

Zero value from converter

The measuring electronics (A/D converter) are not producing a meaningful signal, so the balance cannot calculate weight. This usually points to a problem with the measuring system or the electronics.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the scale off, wait 30 seconds and turn it back on. If the error persists, contact technical service.
  • Do not open the device cover.

Possible causes

  • Measuring system or connection fault
  • A/D converter fault

Do you need service?

A zero value coming from the converter points to the measuring electronics, so the fault must be examined by technical service.

Source: RADWAG official user manual · Open manufacturer document