Adjustment or ınitial mass out of range
According to the manual, the tolerance is 1% for the adjustment weight and 10% for the initial mass. If the weight placed on the platform is outside these limits, the scale does not complete the adjustment or initial mass determination.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the platform completely.
- Use an adjustment weight of the value requested on the display.
- Start the operation again.
Possible causes
- Adjustment weight of the wrong value used
- Another load on the platform during adjustment
- Load on the platform at power-on
Do you need service?
If the procedure repeated with the correct weight and an empty platform still gives Err4, the scale must be adjusted by technical service.
Source: RADWAG official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.