Code on display -Err4- RADWAG C315 scales

Adjustment or ınitial mass out of range

According to the manual, the tolerance is 1% for the adjustment weight and 10% for the initial mass. If the weight placed on the platform is outside these limits, the scale does not complete the adjustment or initial mass determination.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the platform completely.
  2. Use an adjustment weight of the value requested on the display.
  3. Start the operation again.

Possible causes

  • Adjustment weight of the wrong value used
  • Another load on the platform during adjustment
  • Load on the platform at power-on

Do you need service?

If the procedure repeated with the correct weight and an empty platform still gives Err4, the scale must be adjusted by technical service.

Source: RADWAG official user manual · Open manufacturer document