Code on display -Err2- RADWAG C315 scales

Outside zeroing range

The platform scale refuses to zero because the indicator is outside the permitted zeroing range. This limit prevents a real load on the platform from being zeroed by mistake.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the platform completely.
  2. Press the zero key again once the indication is stable.
  3. If you use a container or pallet, tare instead of zeroing.

Possible causes

  • Load on the platform
  • An object touching the platform

Do you need service?

If Err2 continues with the pan empty and correctly installed, the balance's zero point must be checked by technical service.

Source: RADWAG official user manual · Open manufacturer document