Code on display -Err8- RADWAG C315 scales

Tare / zeroing timeout

The balance waits for a stable display for taring and zeroing. If vibration, drafts, or contact with the pan prevent the display from settling, the operation times out and is canceled.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Move the balance to a sturdy table without vibration.
  2. Block drafts from air conditioning, windows, and doors.
  3. Check that the weighing pan is not touching anything and repeat the process.

Possible causes

  • Vibration or air drafts
  • Wobbly table
  • Weighing pan touching something

Do you need service?

If Err8 continues even in a stable environment, the weighing system must be checked by technical service.

Source: RADWAG official user manual · Open manufacturer document