Code on display -LH- RADWAG C315 scales

Initial mass error

At startup the scale compares the measured value with the stored initial mass. According to the manual, LH appears if the indication deviates from the initial mass by more than 10%. A load on the platform at startup is the typical cause.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the scale off and empty the platform completely.
  2. Check that the platform is not touching its surroundings or any cables.
  3. Turn the scale back on.

Possible causes

  • A load on the platform during startup
  • Platform touching something

Do you need service?

If LH persists with an empty, correctly installed pan, the initial mass setting or the measuring system must be checked by technical service.

Source: RADWAG official user manual · Open manufacturer document