Mettler Toledo

ME / newclassic error codes

19 error codes

Review the 19 error and warning messages from the Mettler Toledo ME / NewClassic manufacturer manual, with meanings and fix steps. Device type: precision / laboratory balance.

Covered models: ME, NewClassic ME

NO STABILITY Unable to obtain a stable weighing value The balance cannot reach a stable weighing value. WRONG ADJUSTMENT WEIGHT Wrong calibration weight The wrong weight was used during calibration, or the required weight is not on the weighing pan. REFERENCE TOO SMALL Reference weight too low The reference weight used for piece counting is insufficient. EEPROM ERROR EEPROM memory error An error was detected in the permanent memory of the device. WRONG CELL DATA Weighing cell data error The data belonging to the weighing cell is invalid or faulty. NO STANDARD ADJUSTMENT Standard adjustment data not found NO STANDARD ADJUSTMENT indicates that the standard adjustment (calibration) data was not found on the Mettl… PROGRAM MEMORY DEFECT Program memory error PROGRAM MEMORY DEFECT indicates a program memory error on the Mettler Toledo ME / NewClassic. TEMP SENSOR DEFECT Temperature sensor error TEMP SENSOR DEFECT indicates a temperature sensor error on the Mettler Toledo ME / NewClassic. WRONG LOAD CELL BRAND Incompatible load cell / weighing cell WRONG LOAD CELL BRAND indicates an incompatible load cell / weighing cell on the Mettler Toledo ME / NewCla… WRONG TYPE DATA SET Device type data set error WRONG TYPE DATA SET indicates a device type data set error on the Mettler Toledo ME / NewClassic. OVERLOAD Maximum capacity exceeded The load on the weighing pan exceeds the permitted weighing capacity of the balance. UNDERLOAD Underload / weighing pan warning UNDERLOAD indicates an underload / weighing pan warning on the Mettler Toledo ME / NewClassic. BELOW INITIAL ZERO RANGE Below ınitial zero range BELOW INITIAL ZERO RANGE indicates a below initial zero range condition on the Mettler Toledo ME / NewClassic. BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Backup battery / date and time memory warning The backup power source that preserves the date and time may be depleted. ABOVE INITIAL ZERO RANGE Above ınitial zero range ABOVE INITIAL ZERO RANGE indicates an above initial zero range condition on the Mettler Toledo ME / NewClas… MEM.FULL Memory full The application's record memory is full. The new value cannot be saved. FACTOR OUT OF RANGE Factor out of range The coefficient entered in the multiplier application is not in the valid range. STEP OUT OF RANGE Step value out of range The display step entered is not in the range allowed for the coefficient and resolution. OUT OF RANGE Sample weight outside allowed range In the statistics application, the sample placed is outside the accepted weight range.