Code on display CAL E5 Shimadzu BX-K / BW-K

Wrong calibration weight

The weight placed on the weighing pan does not match the mass value the balance expects for calibration. The process is stopped so that calibration is not done with the wrong reference.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Read the calibration mass value flashing on the display.
  2. Place the correct calibration weight for this value in the center of the weighing pan.
  3. If the weight you have is different, correct the calibration mass value in the menu and repeat the calibration.

Possible causes

  • Using a weight different from the expected value
  • Calibration mass value entered incorrectly in the menu

Do you need service?

This warning clears on its own when the correct weight is used. Service is not required.

Models showing this code: BX-K, BW-K

Source: Shimadzu official user manual · Open manufacturer document