Code on display CAL E2 Shimadzu BX-K / BW-K

Zero point drift during calibration

The zero point drifted more than expected during calibration. In this case the balance cannot take a correct reference point and does not accept the calibration.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan completely.
  2. Check that the weighing pan is not touching anything and is seated correctly.
  3. Wait for the device to warm up and restart the calibration.

Possible causes

  • Load or container left on the weighing pan
  • Weighing pan touching something
  • Change in ambient temperature or insufficient warm-up

Do you need service?

If zero drift continues with an empty and stable weighing pan, service must examine the load cell or electronics.

Models showing this code: BX-K, BW-K