Code on display CAL E3 Shimadzu BX-K / BW-K

PCAL calibration deviation

The value read during the PCAL process (correcting the internal weight value with an external weight) differs too much from what was expected. The process is stopped because a correction made with the wrong weight would corrupt the internal calibration.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Use the correct calibration weight for the value requested on the display.
  2. Place the weight in the center of the weighing pan.
  3. Stabilize the environment and repeat the process.

Possible causes

  • Using an external weight of the wrong value or class
  • Weight not placed in the center of the weighing pan
  • Air drafts or vibration

Do you need service?

If the error continues when PCAL is repeated with the correct weight, service must check the internal weight mechanism.

Models showing this code: BX-K, BW-K

Source: Shimadzu official user manual · Open manufacturer document