PCAL calibration deviation
The value read during the PCAL process (correcting the internal weight value with an external weight) differs too much from what was expected. The process is stopped because a correction made with the wrong weight would corrupt the internal calibration.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Use the correct calibration weight for the value requested on the display.
- Place the weight in the center of the weighing pan.
- Stabilize the environment and repeat the process.
Possible causes
- Using an external weight of the wrong value or class
- Weight not placed in the center of the weighing pan
- Air drafts or vibration
Do you need service?
If the error continues when PCAL is repeated with the correct weight, service must check the internal weight mechanism.
Models showing this code: BX-K, BW-K
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.