Communication command error
The balance received a command code it does not recognize from the computer or printer connection. The balance does not process the command, and the weighing function is not affected.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Compare the command sent with the command list in the manual.
- Make the terminator (delimiter) settings the same on the balance and the computer.
- Match the baud rate, data bit and parity settings on both sides.
Possible causes
- Wrong command code sent
- Terminator (delimiter) setting mismatch
- Different communication settings (baud rate, parity)
Do you need service?
The error is resolved when the communication settings and command code are corrected. Service is not required.
Models showing this code: BX-K, BW-K
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.