Code on display ComErr Shimadzu BX-K / BW-K

Communication command error

The balance received a command code it does not recognize from the computer or printer connection. The balance does not process the command, and the weighing function is not affected.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Compare the command sent with the command list in the manual.
  2. Make the terminator (delimiter) settings the same on the balance and the computer.
  3. Match the baud rate, data bit and parity settings on both sides.

Possible causes

  • Wrong command code sent
  • Terminator (delimiter) setting mismatch
  • Different communication settings (baud rate, parity)

Do you need service?

The error is resolved when the communication settings and command code are corrected. Service is not required.

Models showing this code: BX-K, BW-K

Source: Shimadzu official user manual · Open manufacturer document