Span calibration deviation
The difference between the measured value and the expected calibration mass in span calibration is outside the permitted limit. The balance does not accept the calibration so that it is not adjusted incorrectly.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Use the correct calibration weight for the value flashing on the display.
- Check that there is nothing else on the weighing pan.
- Repeat the calibration from the beginning.
Possible causes
- Using the wrong calibration weight
- Extra load or container on the weighing pan
- Instability caused by the environment
Do you need service?
If span calibration with the correct weight still gives CAL E4, service must check the balance's sensitivity adjustment.
Models showing this code: BX-K, BW-K
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.