Code on display CAL E4 Shimadzu BX-K / BW-K

Span calibration deviation

The difference between the measured value and the expected calibration mass in span calibration is outside the permitted limit. The balance does not accept the calibration so that it is not adjusted incorrectly.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Use the correct calibration weight for the value flashing on the display.
  2. Check that there is nothing else on the weighing pan.
  3. Repeat the calibration from the beginning.

Possible causes

  • Using the wrong calibration weight
  • Extra load or container on the weighing pan
  • Instability caused by the environment

Do you need service?

If span calibration with the correct weight still gives CAL E4, service must check the balance's sensitivity adjustment.

Models showing this code: BX-K, BW-K