Unstable weighing during calibration
Calibration cannot be completed because the reading on the weighing pan cannot be stabilized during span calibration. The manual also lists, on the same line, only one of the internal calibration arms working as a cause. A calibration made with an unstable value would make all weighings incorrect.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Move the balance to a sturdy table away from air drafts and vibration.
- Before calibration, keep the device on for at least 1 hour to warm up.
- Check that the weighing pan is empty and installed correctly, then restart the calibration.
Possible causes
- Air drafts
- Floor or table vibration
- Device not warmed up enough
- Internal calibration arm not working fully
Do you need service?
If CAL E1 still appears on repeated calibration in a stable environment, service must check the internal calibration mechanism.
Models showing this code: BX-K, BW-K
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.