Code on display CAL E1 Shimadzu BX-K / BW-K

Unstable weighing during calibration

Calibration cannot be completed because the reading on the weighing pan cannot be stabilized during span calibration. The manual also lists, on the same line, only one of the internal calibration arms working as a cause. A calibration made with an unstable value would make all weighings incorrect.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Move the balance to a sturdy table away from air drafts and vibration.
  2. Before calibration, keep the device on for at least 1 hour to warm up.
  3. Check that the weighing pan is empty and installed correctly, then restart the calibration.

Possible causes

  • Air drafts
  • Floor or table vibration
  • Device not warmed up enough
  • Internal calibration arm not working fully

Do you need service?

If CAL E1 still appears on repeated calibration in a stable environment, service must check the internal calibration mechanism.

Models showing this code: BX-K, BW-K