Code on display ERR 02 Sartorius Entris II Essential BCE

Zero condition not met for calibration

The device cannot be calibrated because there is a zero point error when calibration is started. For correct calibration, the weighing pan must be empty and the display must show zero at the start.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the sample from the weighing pan.
  2. Zero the device.
  3. If needed, check the preload setting.
  4. Restart the calibration.

Possible causes

  • The device was not zeroed before calibration
  • A sample or load is on the weighing pan
  • The preload setting is not suitable

Do you need service?

Calibration completes normally after the weighing pan is emptied and the device is zeroed. No service is required.

Models showing this code: BCE, Entris II Essential

Source: Sartorius official user manual