Maximum capacity exceeded
HIGH indicates that the load on the weighing pan exceeds the maximum weighing capacity of the balance. An overload invalidates the weighing result and can damage the load cell if it lasts long.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Remove the load from the weighing pan immediately.
- Divide the sample into smaller portions and weigh them.
- Check on the label that the total weight to be weighed (including tare) is below the Max value of the device.
- Do not place a load above capacity on the weighing pan and do not press on the pan.
Possible causes
- A sample above capacity was placed
- An extra container or object was placed on the weighing pan
- Pressure or impact on the weighing pan
Do you need service?
The warning clears when the load is brought below capacity. If it also appears on an empty pan, a service check is required for possible load cell damage.
Models showing this code: BCE, Entris II Essential
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.