Code on display DIS.ERR. Sartorius Entris II Essential BCE

Display format error

The value to be displayed does not fit the display format selected in the menu, so it cannot be shown on the display. This is not a hardware failure but a setting mismatch.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Open the display settings in the menu.
  2. Change the resolution, unit and decimal place settings.
  3. Try displaying the value again.

Possible causes

  • Resolution setting not suitable
  • The selected unit does not suit the value
  • Number of decimal places not suitable

Do you need service?

DIS.ERR. is resolved by correcting the display settings in the menu. No service is required.

Models showing this code: BCE, Entris II Essential

Source: Sartorius official user manual