Code on display ERR 55 Sartorius Entris II Essential BCE

Maximum capacity exceeded

ERR 55 indicates that the load on the weighing pan exceeds the maximum weighing capacity of the balance. An overload invalidates the weighing result and can damage the load cell if it lasts long.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the load from the weighing pan immediately.
  2. Divide the sample into smaller portions and weigh them.
  3. Check on the label that the total weight to be weighed (including tare) is below the Max value of the device.
  • Do not place a load above capacity on the weighing pan and do not press on the pan.

Possible causes

  • A sample above capacity was placed
  • An extra container or object was placed on the weighing pan
  • Pressure or impact on the weighing pan

Do you need service?

The warning clears when the load is brought below capacity. If it also appears on an empty pan, a service check is required for possible load cell damage.

Models showing this code: BCE, Entris II Essential

Source: Sartorius official user manual