Weighing system / weighing pan warning
LOW indicates that the signal of the weighing converter inside the device is too low. It is usually seen when the weighing pan is not installed, or when a load left on the pan at switch-on is removed afterward. Rarely, it points to a weighing system failure.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Seat the weighing pan (and the pan support, if any) correctly in place.
- Switch the device off and on again with the weighing pan empty.
- Check that there is no foreign object under or around the weighing pan.
- Do not open the device.
Possible causes
- The weighing pan is not installed
- A load left on the weighing pan at switch-on was removed afterward
- Weighing system or electronics failure
Do you need service?
If it persists after a restart with the weighing pan installed correctly and empty, a weighing system/electronics failure is likely and technical service is required.
Models showing this code: BCE, Entris II Essential
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.