Code on display LOW Sartorius Entris II Essential BCE

Weighing system / weighing pan warning

LOW indicates that the signal of the weighing converter inside the device is too low. It is usually seen when the weighing pan is not installed, or when a load left on the pan at switch-on is removed afterward. Rarely, it points to a weighing system failure.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Seat the weighing pan (and the pan support, if any) correctly in place.
  2. Switch the device off and on again with the weighing pan empty.
  3. Check that there is no foreign object under or around the weighing pan.
  • Do not open the device.

Possible causes

  • The weighing pan is not installed
  • A load left on the weighing pan at switch-on was removed afterward
  • Weighing system or electronics failure

Do you need service?

If it persists after a restart with the weighing pan installed correctly and empty, a weighing system/electronics failure is likely and technical service is required.

Models showing this code: BCE, Entris II Essential

Source: Sartorius official user manual