Code on display APP.ERR. Sartorius Entris II Essential BCE

Invalid weight for application

The balance measured a weight value that is invalid for the running application. An application (e.g. counting, percent weighing) cannot calculate with a sample that is below the minimum load, negative or missing.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Place the sample on the weighing pan.
  2. Increase the applied load above the minimum load.
  3. If a negative value appears, empty the weighing pan, zero the balance and try again.

Possible causes

  • The sample is lighter than the minimum load
  • Negative weight value
  • No sample was placed on the weighing pan

Do you need service?

APP.ERR. clears by itself when the sample amount is raised above the minimum load or when a sample is placed. No service is required.

Models showing this code: BCE, Entris II Essential

Source: Sartorius official user manual