Code on display -Err Lo- RADWAG C315 scales

Unit weight too light / min > max

In parts counting, a reference that is too light does not give a reliable count. In +/- check mode, a lower limit higher than the upper limit makes the check meaningless.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. In piece counting, use more reference pieces.
  2. In +/- checkweighing mode, check and correct the Min and Max threshold values.

Possible causes

  • Too few pieces placed for the reference
  • Pieces too light
  • Min and Max thresholds entered in reverse

Do you need service?

The message clears once the reference quantity or threshold values are corrected. No service is needed.

Source: RADWAG official user manual · Open manufacturer document