Code on display Err lo KERN & Sohn EWJ

Adjustment weight too small

The EWJ adjusts with an internal weight. If the value read when the internal weight is applied is below the expected range, adjustment is not completed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the pan and check that it is installed correctly.
  2. Check that the transport locking screw is in the released position.
  3. Close the draft shield doors and avoid vibration.
  4. Restart the adjustment after the warm-up time.

Possible causes

  • Load on the pan or pan missing during adjustment
  • Transport lock not released
  • Air drafts or vibration

Do you need service?

If Err lo repeats under proper conditions, the internal adjustment weight mechanism must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document