Code on display Err 19 KERN & Sohn EWJ

Zero point drift

Extra preloads left on the pan or an outdated adjustment shift the zero point. The manual advises removing preloads and adjusting the balance.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove containers and extra parts from the pan.
  2. Switch the balance off and on again with an empty pan.
  3. Start the internal adjustment.

Possible causes

  • Container or extra part left on the pan
  • Adjustment not up to date
  • Change in room temperature

Do you need service?

If Err 19 remains after removing preloads and adjusting, the load cell must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document