Code on display Err 4 KERN & Sohn EWJ

Zero range exceeded at power-on

At power-on the balance takes the pan as zero. If there is too much load on the pan, this zero setting cannot be made.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the pan.
  2. Check that the weighing pan is installed correctly.
  3. Switch the balance off and on again with an empty pan.

Possible causes

  • Sample or container on the pan at power-on
  • Pan installed incorrectly
  • Transport lock not released

Do you need service?

If Err 4 remains at power-on with an empty pan, the load cell must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document