Adjustment weight too large
If the value read when the internal weight is applied is above the expected range, adjustment is not completed. An object left on the pan can cause this.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the pan and check that it is installed correctly.
- Check that the transport locking screw is in the released position.
- Close the draft shield doors and avoid vibration.
- Restart the adjustment after the warm-up time.
Possible causes
- Object on the pan during adjustment
- Pan touched or touching something
- Air drafts or vibration
Do you need service?
If Err hi repeats with an empty pan, the internal weight and weighing system must be checked by a technician.
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.