Measuring signal out of range
The signal from the load cell has gone beyond the limits the converter can read. The manual advises checking the load, the pan and the load cell connection.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Reduce the load on the weighing pan.
- Check that the weighing pan is installed correctly.
- Switch the balance off and on again.
- Do not open the device to check the load cell connection.
Possible causes
- Overload on the pan
- The weighing pan is seated incorrectly
- Load cell connection problem
Do you need service?
If Err 6 remains with correct load and pan, the load cell connection must be checked by a technician.
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.