Code on display Err 6 KERN & Sohn EWJ

Measuring signal out of range

The signal from the load cell has gone beyond the limits the converter can read. The manual advises checking the load, the pan and the load cell connection.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Reduce the load on the weighing pan.
  2. Check that the weighing pan is installed correctly.
  3. Switch the balance off and on again.
  • Do not open the device to check the load cell connection.

Possible causes

  • Overload on the pan
  • The weighing pan is seated incorrectly
  • Load cell connection problem

Do you need service?

If Err 6 remains with correct load and pan, the load cell connection must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document