Code on display 2-Err Vibra AJ / AJH

Calibration difference too large

The balance rejects the new span because it differs too much from the old one. It usually comes from a wrong weight or a wrong procedure.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the nominal value and cleanliness of the weight.
  2. Level the balance using the level indicator.
  3. Close the draft shield and repeat calibration on a stable bench.

Possible causes

  • Wrong or damaged calibration weight
  • Balance not level
  • Air drafts or vibration
  • Mechanism problem

Do you need service?

If a 1% difference remains with a correct weight and setup, the balance must be checked by a technician.

Models showing this code: AJ-2200CEN, AJ-3200CEN, AJ-4200CEN, AJ-6200CEN, AJ-8200CEN, AJ-12KCEN, AJH-2200CEN, AJH-3200CEN, AJH-4200CEN

Source: Vibra official user manual