Code on display o-Err Vibra AJ / AJH

Overload

It appears when the total weight of container and sample is above capacity. During calibration it means the weight used is heavier than the balance capacity.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the weighing pan immediately.
  2. Check the total weight of container and sample.
  3. Tare again and weigh a smaller amount.
  4. For calibration, use a weight that does not exceed capacity.
  • Do not keep using the balance while the error is shown.

Possible causes

  • Container plus sample exceeds capacity
  • Remaining range after tare is too small
  • Calibration weight heavier than capacity
  • Damaged mechanism

Do you need service?

If o-Err appears with a load below capacity, the weighing mechanism must be checked by a technician.

Models showing this code: AJ-2200CEN, AJ-3200CEN, AJ-4200CEN, AJ-6200CEN, AJ-8200CEN, AJ-12KCEN, AJH-2200CEN, AJH-3200CEN, AJH-4200CEN

Source: Vibra official user manual