Code on display 3-Err Vibra AJ / AJH

Zero point off factory setting

It appears at the end of zero point adjustment. The typical example in the manual is starting calibration with an object on the pan.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove everything from the pan.
  2. Check that the weighing pan is seated correctly.
  3. Repeat the span adjustment from the start.

Possible causes

  • Object on the pan when calibration starts
  • Pan not seated correctly
  • Damaged mechanism

Do you need service?

If 3-Err repeats with an empty pan, the internal calibration mechanism must be checked by a technician.

Models showing this code: AJ-2200CEN, AJ-3200CEN, AJ-4200CEN, AJ-6200CEN, AJ-8200CEN, AJ-12KCEN, AJH-2200CEN, AJH-3200CEN, AJH-4200CEN

Source: Vibra official user manual