Code on display 1-Err Vibra AJ / AJH

Calibration weight too light

On AJ-CEN models the span weight must be at least 50% of capacity. Calibration is not carried out while this error is shown.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Use a calibration weight of at least half the capacity.
  2. Choose a weight close to capacity if possible.
  3. Repeat the calibration from the beginning.

Possible causes

  • Weight lighter than half of capacity
  • Wrong nominal weight selected

Do you need service?

If the error remains with a correct weight, the balance adjustment must be checked by a technician.

Models showing this code: AJ-2200CEN, AJ-3200CEN, AJ-4200CEN, AJ-6200CEN, AJ-8200CEN, AJ-12KCEN, AJH-2200CEN, AJH-3200CEN, AJH-4200CEN