Instability during calibration
Calibration cannot be completed because the reading cannot be stabilized during balance calibration. A calibration made in an unstable environment also affects later moisture measurements.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Place the device on a level, sturdy surface.
- Move it away from sources of vibration and air drafts.
- Repeat the calibration.
Possible causes
- Uneven or wobbly surface
- Vibration
- Air drafts or breeze
Do you need service?
If Er502 continues even in a stable environment, service must check the weighing unit.
Models showing this code: MOC-120H
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.