Code on display Er502 Shimadzu MOC-120H

Instability during calibration

Calibration cannot be completed because the reading cannot be stabilized during balance calibration. A calibration made in an unstable environment also affects later moisture measurements.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Place the device on a level, sturdy surface.
  2. Move it away from sources of vibration and air drafts.
  3. Repeat the calibration.

Possible causes

  • Uneven or wobbly surface
  • Vibration
  • Air drafts or breeze

Do you need service?

If Er502 continues even in a stable environment, service must check the weighing unit.

Models showing this code: MOC-120H