Code on display Er103 Shimadzu MOC-120H

Sample too light

The sample weight is below 0.5 g, the lowest value the MOC-120H can measure. A reliable moisture result cannot be obtained with so little sample.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Press the TARE/RESET key to clear the error message.
  2. Place at least 0.5 g of sample and restart the measurement.

Possible causes

  • Very little sample placed on the weighing pan
  • Measurement started without placing the sample on the weighing pan

Do you need service?

The error does not occur when measuring with enough sample. Service is not required.

Models showing this code: MOC-120H

Source: Shimadzu official user manual · Open manufacturer document