Invalid moisture measurement / weight ıncrease
The weight increased by more than 0.1 g during the measurement. According to the manual, this is usually seen when sample is added while the measurement is running. A weight increase during drying invalidates the moisture calculation.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Press the TARE/RESET key to clear the error message.
- Restart the measurement from the beginning with a new sample.
- Do not add sample or touch the weighing pan while the measurement is running.
Possible causes
- Adding sample to the weighing pan during measurement
- Touching the weighing pan or the sample
Do you need service?
The error does not occur when the measurement is repeated under correct conditions. Service is not required.
Models showing this code: MOC-120H
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.