Code on display Er201 Shimadzu MOC-120H

Invalid moisture measurement / weight ıncrease

The weight increased by more than 0.1 g during the measurement. According to the manual, this is usually seen when sample is added while the measurement is running. A weight increase during drying invalidates the moisture calculation.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Press the TARE/RESET key to clear the error message.
  2. Restart the measurement from the beginning with a new sample.
  • Do not add sample or touch the weighing pan while the measurement is running.

Possible causes

  • Adding sample to the weighing pan during measurement
  • Touching the weighing pan or the sample

Do you need service?

The error does not occur when the measurement is repeated under correct conditions. Service is not required.

Models showing this code: MOC-120H

Source: Shimadzu official user manual · Open manufacturer document