Code on display Er401 Shimadzu MOC-120H

Weighing unit communication error

Communication between the device's control unit and the weighing unit has been lost. Weight cannot be read and no measurement can be made in this case.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off, wait a few seconds and turn it on again.

Possible causes

  • Temporary electronic communication error
  • Internal connection or board fault

Do you need service?

If Er401 continues after restarting, service must examine the internal connection or electronic board.

Models showing this code: MOC-120H

Source: Shimadzu official user manual · Open manufacturer document