Code on display Er202 Shimadzu MOC-120H

Invalid moisture measurement / measured weight too low

The measured weight dropped below -1 g, so the moisture value is invalid. Weight being removed from the weighing pan during the measurement corrupts the calculation.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press the TARE/RESET key to clear the error message.
  2. Check that the weighing pan is seated correctly.
  • Restart the measurement from the beginning and do not touch the weighing pan during it.

Possible causes

  • Sample removed or spilled from the weighing pan during measurement
  • Weighing pan shifted out of place

Do you need service?

If Er202 also appears on measurements repeated under correct conditions, service must check the weighing unit.

Models showing this code: MOC-120H

Source: Shimadzu official user manual · Open manufacturer document