Code on display Er501 Shimadzu MOC-120H

Invalid calibration weight

The weight used in balance calibration does not match the calibration value the device expects. The process is stopped so that calibration is not done with the wrong reference.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check that there is nothing else on the weighing pan.
  2. Repeat the calibration using a calibration weight of the correct value.

Possible causes

  • Using a calibration weight of the wrong value
  • Extra load on the weighing pan

Do you need service?

The error clears when calibration is done with the correct weight. Service is not required.

Models showing this code: MOC-120H