Code on display Err i0 Shimadzu BX-K / BW-K

PCAL password error

The password entered to access the PCAL (internal weight correction) process is wrong. After the error, '___' appears on the display and the process is not carried out.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check the password.
  2. Press the relevant key to return the balance to the mass display.
  3. Restart the PCAL process with the correct password.

Possible causes

  • Wrong password entered
  • Password was changed

Do you need service?

The process continues when the correct password is entered. If the password is unknown, contact the balance administrator.

Models showing this code: BX-K, BW-K

Source: Shimadzu official user manual · Open manufacturer document