Code on display Err 24 Shimadzu BX-K / BW-K

Supply voltage abnormal

The setting cannot be saved because the supply voltage is not normal. Changes are not permanent until the voltage is corrected.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that you are using the original AC adapter supplied with the device.
  2. Plug the adapter directly into the outlet, not an extension cord or power strip.
  3. Check the connections and save the setting again.

Possible causes

  • Unsuitable AC adapter
  • Low or fluctuating mains voltage
  • Loose power connection

Do you need service?

If Err 24 continues even with a suitable power supply, service must check the balance's power circuit.

Models showing this code: BX-K, BW-K

Source: Shimadzu official user manual · Open manufacturer document