Code on display SYS.ERR. Sartorius Entris II Essential BCE

System data / memory error

The system data is faulty and a memory error was detected in the control module. The adjustment and calibration data may not be reliable.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Switch the device off and on again.
  2. If the error has cleared, check the settings and calibration.

Possible causes

  • Memory error in the control module
  • Power outage during operation
  • Electronics board failure

Do you need service?

If SYS.ERR. recurs despite a restart, technical service is required for the control module memory.

Models showing this code: BCE, Entris II Essential

Source: Sartorius official user manual