Print function unavailable
The [Print] key is locked. Because the data interface is set to xBPI mode for print output, a printout cannot be made with the key.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check the data interface / communication settings in the menu.
- Reset the menu to factory settings.
- Try printing again.
Possible causes
- The data interface is set to xBPI mode
- The communication settings in the menu were changed
Do you need service?
If the [Print] key stays locked after returning to factory settings, contact technical service for a software/interface problem.
Models showing this code: BCE, Entris II Essential
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.