Internal communication error
The device cannot get a weight value because communication between the control module and the weighing cell has been interrupted. No weighing result can be displayed during this time.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Wait a few seconds for the control module to reestablish communication.
- If it does not clear, switch the device off and back on after 30 seconds.
- Do not open the device or tamper with the internal cables.
Possible causes
- Temporary loss of communication between the control module and the weighing cell
- Electrical interference or power fluctuation
- Internal connection or electronics failure
Do you need service?
If the communication error recurs, the connection between the control module and the weighing cell must be checked by technical service.
Models showing this code: BCE, Entris II Essential
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.