Code on display HIGH Sartorius Cubis II MCE

Maximum capacity exceeded

The load on the pan exceeds the maximum capacity of the device. Overload invalidates the result and may damage the weighing system.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the load from the weighing pan immediately.
  2. Weigh the sample in smaller portions.
  • Do not place a load above capacity on the weighing pan.

Possible causes

  • Sample above capacity
  • Extra object placed on the pan
  • Pressing down on the pan

Do you need service?

The warning clears once the load falls below capacity. If it persists with an empty pan, a service check is required.

Models showing this code: MCE, Cubis II, Cubis MCE

Source: Sartorius official user manual